Multi Parameter Patient Monitor 

Features:

12.1" Colour TFT LCD display with wave forms upto 8 Channels Portable streamline handle design Monitor measures from Adult to Neonates Graphic and tabular Trend for 96 hours for all parameters Multi Screen format Audible and Visual alarms with adjustable alarm ranges Powerful data management and storage capacity Manual or Automatic NIBP measuring with 500 unit memory Rechargeable battery backup Weight : 4.5Kg approx. 

Respiration 

Method: RA-LL impedance 

Measurement range: Adult: 0-120rpm, paediatric/neonatal: 

  • 0-150rpm
  • Apnea alarm: YES

Temperature

  • Resolution: 0.10C
  • Channel: Dual-channel
  • Measurement range: 0-50 0C

IBP (Ungradable)

  • Channel: 2 channels
  • Pressure transducer: Sensitivity: 5( mVN/mmHg) 
  • Transducer sites: ART, PA, CVP, RAP, LAP, ICP
  • Alarm range: -10~300mmHg

ECG

  • 5-lead or 12-lead selectable
  • Input: 5-lead: RA; LA; RL; LL; V or R; L; N; F; C 
  • Lead selection: I; II; III; avR; avL; avF; Vx; CAL 
  • Gain selection: x0.25,x0.5,x I,x 2,auto
  • Sweep speed: 12.5mm/s, 25mm/s, 50mm/s 
  • Accuracy:  1bpm or 1 %, whichever is greater
  • Protection: Withstand 4000V AC/50Hz voltage in isolation against electrosurgical and defibrillation Calibration signal: lmV5%
  • Operation modes: Diagnostic, Monitor, Surgery 
  • Heart rate range: Adult: 15-300bpm Paediatric/Neonatal: 15- 350bpm
  • Measurement range: -2.0mV~2.0mV
  • S-T detection: YES, Arrhythmia analysis: YES 
  • Alarm: YES, audible and visual alarm, alarm events recallable

NIBP

  • Method: Oscillometry
  • Operation modes: Manual/ Automatic/STAT 
  • Measurement unit: mmHg/KPa selectable 
  • Measurement types: Systolic, Diastolic, Mean 
  • Measurement range: Adult: 10-270mmHg 
  • Over-pressure protection: YES
  • Resolution: lmmHg
  • Accuracy: The mean error shall be less than 5 mmHg, the standard deviation shall be less than 8mmHg
  • Alarm: Systolic, Diastolic, Mean

Sp02 

  • Measurement range: 0-100% 
  • Accuracy: 2%(70~100%); 0~69% 
  • unspecified 
  • Alarm range: 0-100% 